Show simple item record

dc.contributor.authorGregory, RA
dc.contributor.authorMurray, I
dc.contributor.authorGear, J
dc.contributor.authorAldridge, MD
dc.contributor.authorLevine, D
dc.contributor.authorFowkes, L
dc.contributor.authorWaddington, WA
dc.contributor.authorChua, S
dc.contributor.authorFlux, G
dc.date.accessioned2017-02-07T14:52:53Z
dc.date.issued2017-01-07
dc.identifier.citationPhysics in medicine and biology, 2017, 62 (1), pp. 17 - 30
dc.identifier.issn0031-9155
dc.identifier.urihttps://repository.icr.ac.uk/handle/internal/407
dc.identifier.eissn1361-6560
dc.identifier.doi10.1088/1361-6560/62/1/17
dc.description.abstractIodine-123 mIBG imaging is widely regarded as a gold standard for diagnostic studies of neuroblastoma and adult neuroendocrine cancer although the optimal collimator for tumour imaging remains undetermined. Low-energy (LE) high-resolution (HR) collimators provide superior spatial resolution. However due to septal penetration of high-energy photons these provide poorer contrast than medium-energy (ME) general-purpose (GP) collimators. LEGP collimators improve count sensitivity. The aim of this study was to objectively compare the lesion detection efficiency of each collimator to determine the optimal collimator for diagnostic imaging. The septal penetration and sensitivity of each collimator was assessed. Planar images of the patient abdomen were simulated with static scans of a Liqui-Phil™ anthropomorphic phantom with lesion-shaped inserts, acquired with LE and ME collimators on 3 different manufacturers' gamma camera systems (Skylight (Philips), Intevo (Siemens) and Discovery (GE)). Two-hundred normal and 200 single-lesion abnormal images were created for each collimator. A channelized Hotelling observer (CHO) was developed and validated to score the images for the likelihood of an abnormality. The areas under receiver-operator characteristic (ROC) curves, Az, created from the scores were used to quantify lesion detectability. The CHO ROC curves for the LEHR collimators were inferior to the GP curves for all cameras. The LEHR collimators resulted in statistically significantly smaller Azs (p  <  0.05), of on average 0.891  ±  0.004, than for the MEGP collimators, 0.933  ±  0.004. In conclusion, the reduced background provided by MEGP collimators improved 123I mIBG image lesion detectability over LEHR collimators that provided better spatial resolution.
dc.formatPrint-Electronic
dc.format.extent17 - 30
dc.languageeng
dc.language.isoeng
dc.publisherIOP PUBLISHING LTD
dc.rights.urihttps://creativecommons.org/licenses/by/4.0
dc.subjectHumans
dc.subjectNeuroblastoma
dc.subject3-Iodobenzylguanidine
dc.subjectRadionuclide Imaging
dc.subjectROC Curve
dc.subjectPhantoms, Imaging
dc.subjectPhotons
dc.subjectChild
dc.titleObjective comparison of lesion detectability in low and medium-energy collimator iodine-123 mIBG images using a channelized Hotelling observer.
dc.typeJournal Article
rioxxterms.versionofrecord10.1088/1361-6560/62/1/17
rioxxterms.licenseref.urihttps://creativecommons.org/licenses/by/4.0
rioxxterms.licenseref.startdate2017-01
rioxxterms.typeJournal Article/Review
dc.relation.isPartOfPhysics in medicine and biology
pubs.issue1
pubs.notesNo embargo
pubs.organisational-group/ICR
pubs.organisational-group/ICR/Primary Group
pubs.organisational-group/ICR/Primary Group/ICR Divisions
pubs.organisational-group/ICR/Primary Group/ICR Divisions/Radiotherapy and Imaging
pubs.organisational-group/ICR/Primary Group/ICR Divisions/Radiotherapy and Imaging/Radioisotope Physics
pubs.organisational-group/ICR/Primary Group/ICR Divisions/Radiotherapy and Imaging/Radioisotope Physics/Radioisotope Physics (hon.)
pubs.organisational-group/ICR/Primary Group/Royal Marsden Clinical Units
pubs.organisational-group/ICR
pubs.organisational-group/ICR/Primary Group
pubs.organisational-group/ICR/Primary Group/ICR Divisions
pubs.organisational-group/ICR/Primary Group/ICR Divisions/Radiotherapy and Imaging
pubs.organisational-group/ICR/Primary Group/ICR Divisions/Radiotherapy and Imaging/Radioisotope Physics
pubs.organisational-group/ICR/Primary Group/ICR Divisions/Radiotherapy and Imaging/Radioisotope Physics/Radioisotope Physics (hon.)
pubs.organisational-group/ICR/Primary Group/Royal Marsden Clinical Units
pubs.publication-statusPublished
pubs.volume62
pubs.embargo.termsNo embargo
icr.researchteamRadioisotope Physics
dc.contributor.icrauthorGear, Jonathan


Files in this item

Thumbnail

This item appears in the following collection(s)

Show simple item record

https://creativecommons.org/licenses/by/4.0
Except where otherwise noted, this item's license is described as https://creativecommons.org/licenses/by/4.0