Lung stereotactic body radiotherapy with an MR-linac - Quantifying the impact of the magnetic field and real-time tumor tracking.
(ELSEVIER IRELAND LTD, 2016-06-01)
BACKGROUND AND PURPOSE: There are concerns that radiotherapy doses delivered in a magnetic field might be distorted due to the Lorentz force deflecting secondary electrons. This study investigates this effect on lung ...
Using dual-energy x-ray imaging to enhance automated lung tumor tracking during real-time adaptive radiotherapy.
(WILEY, 2015-12-01)
PURPOSE: Real-time, markerless localization of lung tumors with kV imaging is often inhibited by ribs obscuring the tumor and poor soft-tissue contrast. This study investigates the use of dual-energy imaging, which can ...
Treating locally advanced lung cancer with a 1.5T MR-Linac - Effects of the magnetic field and irradiation geometry on conventionally fractionated and isotoxic dose-escalated radiotherapy.
(ELSEVIER IRELAND LTD, 2017-11-01)
PURPOSE: This study investigates the feasibility and potential benefits of radiotherapy with a 1.5T MR-Linac for locally advanced non-small cell lung cancer (LA NSCLC) patients. MATERIAL AND METHODS: Ten patients with LA ...
Geometric and dosimetric evaluations of atlas-based segmentation methods of MR images in the head and neck region.
(IOP Publishing, 2018-07-11)
Owing to its excellent soft-tissue contrast, magnetic resonance (MR) imaging has found an increased application in radiation therapy (RT). By harnessing these properties for treatment planning, automated segmentation methods ...