Joint radial trajectory correction for accelerated T2 * mapping on an MR-Linac.
Date
2023-05-27ICR Author
Author
Bano, W
Holmes, W
Goodburn, R
Golbabaee, M
Gupta, A
Withey, S
Tree, A
Oelfke, U
Wetscherek, A
Type
Journal Article
Metadata
Show full item recordAbstract
BACKGROUND: T2 * mapping can characterize tumor hypoxia, which may be associated with resistance to therapy. Acquiring T2 * maps during MR-guided radiotherapy could inform treatment adaptation by, for example, escalating the dose to resistant sub-volumes. PURPOSE: The purpose of this work is to demonstrate the feasibility of the accelerated T2 * mapping technique using model-based image reconstruction with integrated trajectory auto-correction (TrACR) for MR-guided radiotherapy on an MR-Linear accelerator (MR-Linac). MATERIALS AND METHODS: The proposed method was validated in a numerical phantom, where two T2 * mapping approaches (sequential and joint) were compared for different noise levels (0,0.1,0.5,1) and gradient delays ([1, -1] and [1, -2] in units of dwell time for x- and y-axis, respectively). Fully sampled k-space was retrospectively undersampled using two different undersampling patterns. Root mean square errors (RMSEs) were calculated between reconstructed T2 * maps and ground truth. In vivo data was acquired twice weekly in one prostate and one head and neck cancer patient undergoing treatment on a 1.5 T MR-Linac. Data were retrospectively undersampled and T2 * maps reconstructed, with and without trajectory corrections were compared. RESULTS: Numerical simulations demonstrated that, for all noise levels, T2 * maps reconstructed with a joint approach demonstrated less error compared to an uncorrected and sequential approach. For a noise level of 0.1, uniform undersampling and gradient delay [1, -1] (in units of dwell time for x- and y-axis, respectively), RMSEs for sequential and joint approaches were 13.01 and 9.32 ms, respectively, which reduced to 10.92 and 5.89 ms for a gradient delay of [1, 2]. Similarly, for alternate undersampling and gradient delay [1, -1], RMSEs for sequential and joint approaches were 9.80 and 8.90 ms, respectively, which reduced to 9.10 and 5.40 ms for gradient delay [1, 2]. For in vivo data, T2 * maps reconstructed with our proposed approach resulted in less artifacts and improved visual appearance compared to the uncorrected approach. For both prostate and head and neck cancer patients, T2 * maps reconstructed from different treatment fractions showed changes within the planning target volume (PTV). CONCLUSION: Using the proposed approach, a retrospective data-driven gradient delay correction can be performed, which is particularly relevant for hybrid devices, where full information on the machine configuration is not available for image reconstruction. T2 * maps were acquired in under 5 min and can be integrated into MR-guided radiotherapy treatment workflows, which minimizes patient burden and leaves time for additional imaging for online adaptive radiotherapy on an MR-Linac.
Collections
Subject
MR-Linac
T2* mapping
gradient delay correction
hypoxia imaging
Research team
Radiother Phys Modelling
Language
eng
Date accepted
2023-04-28
License start date
2023-05-27
Citation
Medical Physics, 2023,
Publisher
WILEY